Meditech Debrillator Monitor can work both in Automated External Debrillator (AED) mode and manuell mode and record parameters of ECG, HR and SpO2
Meditech Debrillator Monitor can work both in Automated External Debrillator (AED) mode and manuell mode and record parameters of ECG, HR and SpO2. Merkmale: ·
Works both in automated external debrillator (AED) and manuell modes · Records ECG, HR and SpO2 · It has voice and visual prompts for charge prompt and shockable rhythm · It has a high-resolution Anzeige and a thermal printer
| Anzeige: | 7 Bildschirm(17.7cm)diagonally |
| Sweep speed: | 25mm/ sec |
| Number of leads for ECG: | 3-lead ECG cable or 5-lead ECG cable, paddles |
| Output: | Biphasic |
| Energy sequence: | Monitor Anzeige indicates both selected and delivered energy |
| Charge time: | <7sec with a new fully charged Akku |
| Charge prompts: | Voice and visual prompts |
| Electrode impedance measurement Spektrum: | 0-250Ω |
| Lead selection: | Displayed on Monitor, paddles, I, II, III, AVR, AVL, AVF, V |
| ECG size: | 0.25, 0.5, 1, 1.5, 2, 4 cm/mV Anzeige on Monitor |
| Heart rate: | 20-300bpm |
| Heart rate alarm: | On/Off displayed on Monitor, user-selectable |
| Smart alarm: | Beeper/voice prompts indicate shockable rhythm |
| SpO2 measurement Spektrum: | 30~100%, ±2% between 80%~90%, others ±5% |
| Alarm Spektrum for SpO2 measurement Spektrum: | User set high limit and low limit |
| Alarm accuracy for SpO2 measurement Spektrum: | ±10% within setting values |
| Alarm time accuracy for SpO2 measurement Spektrum: | Less than 12secs |
| Funktionen of AED mode: | Auto analyse and charge x3 with programmable auto energy level selection, Bildschirm prompts, and voice prompts |
| Shockable rhythms of AED mode: | Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm QRS complex wave duration >=140ms |
| Charge control of AED mode: | Control on device front panel, press key on paddle |
| Prompts of AED mode: | Voice and visual prompts |
| Energy selection of manuell mode: | Selectable at 2, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200, 300, 360 joules |
| Synchronization of manuell mode: | Synchronizes Defibrillator pulse to patient’s R-wave. ”SYNC” Nachricht displayed on Monitor |
| Printer paper: | 50mm thermal paper |
| Paper speed: | 12.5mm/sec, 25mm/sec, 50mm/sec. User-selectable 6-second delay |
| Printing method: | High-resolution, thermal print head |
| Print-out modes: | manuell or automatic, user-configurable |
| On/Off control: | Front panel and paddle |
| Automatic Funktion: | 9-seconds recording initiated by alarm activation or Defibrillator charge or Defibrillator discharge |
| Akku: | Rechargeable, Ni-MH Akku, 12V |
| Akku life: | For a new, fully charged Akku- 60 Defibrillator discharges at maximum energy, or 3 hours minimum of kontinuierlich ECG Überwachung. Additional parameters will effect operating time with different Funktionen |